Medium throughput system for recording from 64 electrodes on 4 or 8 samples

MED Connector
(MED-C03)

The MED Connector holds and connects the MED Probe (MEA) to the amplifiers. Thanks to the MED Probe’s lowest-impedance electrodes the MED Connector holding the MED Probe is connected to the amplifier with 2m-long cable, that does not generate signal attenuation. This unique MED64 system configuration will provide the great flexibilities for your experiments.

Features

  • No active onboard electrical circuitry; Connectors can safely be placed in a humidified environment (e.g. cell culture incubator).
  • Provides clear optical path through MED Probe for microscopy.
  • Provides a mechanically-stable platform for the MED Probe.

Specifications

MED Connector (MED-C03)
MED Probe securing mechanism Screw down
Material Aluminum, Gold plate for contact pins
Contact resistance < 30 mΩ
Printed wiring 4 layer
– 1st, 4th : ground,
– 2nd, 3rd : signal line
Weight 480 g
Dimensions W174 x L150 x H21 (mm)

 

MED ThermoConnector
(MED- CP04)

Features

  • Heater and thermocouple sensors are incorporated into the bottom unit of the MED Connector.
  • Bath temperature in the MED Probe is maintained using a dedicated temperature controller.
  • Can be used with or without perfusion. Included tubing through connector provides in-line solution heating.
  • Aperture in the unit bottom allows observing biological samples with microscope.

Specifications

MED ThemoConnector (MED-CP04)
Connector Unit
MED Probe securing mechanism Screw down
Material Aluminum, Gold plate for contact pins
Contact resistance < 30 mΩ
Printed circuit board 4 layers
– 1st, 4th : ground,
– 2nd, 3rd : signal line
Heating device Transistors (2)
Temperature accuracy <+/- 1°C (for solution in the MED Probe)
Range of verified <1°C accuracy 32°C – 37°C (RT=25°C)
Weight 510 g
Dimensions W200 x D20 x D105 (mm)
Temperature Controller
Temperature control unit OMRON E5GC
Controlling method Auto-tuning PID temperature control
Resolution 0.1°C
Power supply voltage DC ±12V
Consumption current +1A, -0.1A (typ.) / +2A (max.)
Weight 350 g
Dimensions W150 x D110 x H50 (mm)
Power supply unit, Input AC 100-240 V (50-60 Hz)
Power supply unit, Output DC ±12V

 